Tuning and Calibrating
Running Auto Tune and Calibration
66
TSQ Quantum XLS Series User Guide
Thermo Scientific
Figure 36.
Typical parameter optimization plot
6. To accept the results of the tuning and calibration procedure, click
Accept
.
After you accept the results of the tuning and calibration procedure, a message box
prompts you to copy the positive ion tuning and calibration settings to the negative ion
mode.
• If you have already tuned and calibrated the instrument successfully in the negative
ion mode, click
No
. (Do not copy the positive ion mode parameters to the negative
ion mode.)
• If you have not tuned and calibrated the instrument in the negative ion mode,
click
Yes
.
Note
If you intend to perform high-sensitivity, negative-ion mass spectrum
analysis, Thermo Fisher Scientific recommends that you also perform a full tune
and calibration of the instrument in the negative ion mode.