System Shutdown, Startup, and Reset
Starting Up the System after a Complete Shutdown
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
55
7. Turn off the GC. Follow the procedure described in the manual that came with the GC.
8. Turn off the autosampler by using the main power on/off switch.
9. Press the System Power Off button (
) to remove all power from the TSQ
Quantum XLS or TSQ Quantum GC system.
10. If the system is not to be used for an extended period, turn off the argon collision gas
supply at the tank.
11. Turn off the data system:
a. Choose
Start > Shut Down
from the Windows
®
task bar. The Shut Down Windows
dialog box appears.
b. To start the Windows shutdown procedure, select
Shut down
and click
OK
.
12. Turn off the (optional) printer by using the on/off switch.
Starting Up the System after a Complete Shutdown
Starting the TSQ Quantum XLS or TSQ Quantum GC system after it has been shut down
completely requires the following:
•
Restoring Power to the TSQ Quantum XLS or TSQ Quantum GC system
•
•
•
Starting Up the Mass Spectrometer
•
•
Setting Up Conditions for Operation
Note
If you plan to perform routine or preventive system maintenance on the mass
spectrometer only, you can leave the argon, data system, GC, and autosampler on. In
this case, the shutdown procedure is complete. However, if you do not plan to operate
your system for an extended period of time, Thermo Fisher Scientific recommends
that you turn off the GC, data system, and autosampler as described in steps 7
through 12 below.
CAUTION
Burn Hazard
. The ion source, GC oven, injector, and transfer line are hot.
Allow them to cool to room temperature before touching them.