System Shutdown, Startup, and Reset
Placing the System in Standby Mode
52
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
Figure 27.
System Power Off button on the front of the electronics module
Figure 28.
Power panel on the rear of the electronics module
Placing the System in Standby Mode
If you are not going to use the TSQ Quantum XLS or TSQ Quantum GC system for a short
period of time, such as overnight or over weekends, it does not need to be shut down
completely. When you are not going to operate the system for 12 hours or more, you can leave
the system in standby mode.
Y
To place the TSQ Quantum XLS or TSQ Quantum GC system in standby mode
1. Wait until data acquisition, if any, is complete.
2. From the TSQ EZ Tune window, choose
Control > Standby
(or click the
On/Standby
button) to put the mass spectrometer in standby
.
System Power
Off button
!
!