6
Maintenance
Adding Calibration Compound
118
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
Figure 60.
Location of the calibration compound vial compartment
4. Remove the cover to the calibration compound vial compartment.
5. Remove the calibration compound vial from the calibration gas flow module. See
a. Use a 9/16 in. wrench to remove the nut from the calibration compound vial.
b. Slide the vial away from the calibration valve fitting with the ferrule and nut attached.
c.
Remove the ferrule and nut.
6. Add the calibration compound.
a. Draw up no more than 0.1 mL of calibration compound into a clean syringe.
b. Insert the syringe in the calibration compound vial until it comes in contact with the
white frit at the bottom of the vial.
c.
Inject no more than 0.1 mL of calibration compound into the white frit. The frit
absorbs the calibration compound, so if the calibration compound is pooled
(calibrant covers the top of the frit), then you have too much in the vial. Immediately
pour out the excess calibrant according to local environmental regulations.
Calibration compound vial
compartment
CAUTION
Having more than 0.1 mL of the calibration compound can damage
the calibration gas flow module. Be sure that there is less than 0.1 mL of the
calibration compound in the vial.