Maintenance
Disassembling and Reassembling the TSQ Quantum GC Ion Source Completely
110
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
Disassembling and Reassembling the TSQ Quantum GC Ion Source
Completely
Disassemble the ion source completely to clean the ion source block or replace the EI/CI
Source PCB.
See
for the location of the ion source components.
Y
To disassemble the ion source completely
1. Prepare a clean work area by covering the area with lint-free cloth.
2. Shut down and vent the TSQ Quantum XLS. See
3. Wearing clean, lint- and powder- free gloves, remove the ion source. See
4. Remove the ion source from the magnet yoke.
Tools Needed
• Gloves, cleanroom grade (P/N 23827-0008 and 23827-0009)
• Lint-free cloth
Frequency
As needed
CAUTION
Shock Hazard
. Unplug the TSQ Quantum XLS before proceeding.