10. Close the lid of the ion source vacuum manifold.
Removing the Ion Source Lens Assembly
The ion source lens assembly includes lens L1, L2, L3, and L4. You must remove and
disassemble the ion source lens assembly to clean the lenses. A retainer clip fastens the lens
assembly to the ion source block.
To remove the ion source lens assembly
1. Wearing clean, lint- and powder- free gloves, unscrew the threaded hexagonal standoff
and remove the retainer clip that secures the lens assembly to the ion source block.
See
.
2. Pull the lens assembly out of the ion source.
3. Place the ion source and lens assembly on a clean surface.
Figure 52.
Lens assembly removed from the ion source
Disassembling the Ion Source Lens Assembly
Y
To disassemble the lens assembly
1. Pull the L1, L2, L3 lens assembly from the L4 lens assembly.
Ion source
Retainer clip
Threaded
standoff
Lens
assembly