9
Shutting the MS System Down Completely
Turning off the LC, Gases, Data System, and Autosampler
34
TSQ Altis, Quantis, and Fortis Hardware Manual
Thermo Scientific
Figure 20.
Electronic service switch and main power switch
3. Unplug the MS’s power cord from the electrical outlet.
If you plan to have the system off for an extended period of time, Thermo Fisher Scientific
recommends that you also turn off the LC, gases, data system, and autosampler.
Turning off the LC, Gases, Data System, and Autosampler
1. If the LC system is included, turn it off as described in the LC manual.
2. Turn off the argon and nitrogen gas supplies at their tanks.
3. Shut down the data system computer and turn off the monitor.
4. If an autosampler and printer are included, use their On/Off switches to turn them off.
No.
Description
1.
Electronics service switch- service mode
2.
Main power switch-off
2
1
CAUTION
Do not disconnect the power supply cord from the MS while the other
end is still plugged into the electrical outlet.
Tip
When performing routine or preventive system maintenance on the MS, you do
not need to turn off the LC, gases, data system, and autosampler. In this case, the
shutdown procedure is complete.