Advanced Troubleshooting
Convectron Gauge and Foreline Adapter with Hose
416
TSQ 9000 Hardware Manual
Thermo Scientific
Convectron Gauge and Foreline Adapter with Hose
A convectron gauge measures and controls the pressure in the foreline that is connected to the
turbomolecular pump and rough pump. If the foreline pressure gets too high, convectron
gauge decreases the power going to the turbomolecular pump. The convectron gauge and
foreline adapter are connected to the foreline hose, which is connected to the turbomolecular
pump.
To replace a convectron gauge and foreline adapter with hose
1. Shut down your system by following the steps in
“Shutting Down the TSQ 9000 System”
. You do not have to remove the ion source cartridge.
2. Slide the top cover panel toward the back of the instrument and lift it off.
3. Use a T20 Torxhead screwdriver to remove the four screws securing the left hand front
panel to the chassis. One of the screws is located on the interior front panel of the
instrument.
Figure 442.
Locating the Left Hand Front Panel Screws
4. Lift the left hand front panel off the instrument.
5. Move the GC away from the TSQ 9000 instrument.
Note
See the
ISQ and TSQ GC-MS Spare Parts Guide
for ordering information.
Screws
Left Hand Front
Panel
Screw on Front
of Instrument