Advanced Troubleshooting
Replacing Fans and Filters
394
TSQ 9000 Hardware Manual
Thermo Scientific
Chassis Cooling Fan and Fan Plenum
To replace the chassis cooling fan and fan plenum
1. Shut down your system by following the steps in
“Shutting Down the TSQ 9000 System”
.
2. Use a T20 Torxhead screwdriver to loosen the top cover panel screw from the front panel.
If you are using hydrogen you must also remove the hydrogen safety screws. Slide the top
cover panel toward the back of the instrument and lift it off.
3. Remove the four screws securing the left hand front panel to the chassis. One of the
screws is located on the interior front panel of the instrument.
Figure 419.
Locating the Left Hand Front Panel Screws
4. Lift the left hand front panel off the instrument.
Note
See the
ISQ and TSQ GC-MS Spare Parts Guide
for information about ordering this
component.
Note
You do not have to remove the ion source cartridge or column or vent the
instrument. If you are using hydrogen as a carrier gas, power off the GC.
Screws
Left Hand Front
Panel
Screw on Front
of Instrument