Advanced Troubleshooting
Replacing the Power Supplies
358
TSQ 9000 Hardware Manual
Thermo Scientific
Power Module
To replace the power module
1. Shut down your system by following the steps in
“Shutting Down the TSQ 9000 System”
.
2. Slide the top cover panel toward the back of the instrument and lift it off.
3. Disconnect the power cables from the side of the instrument.
Note
See the
ISQ and TSQ GC-MS Spare Parts Guide
for information about ordering this
component.
Note
You do not have to remove the ion source cartridge or the column or vent the
instrument. If you are using hydrogen as a carrier gas, power off the GC.