3
Advanced Troubleshooting
Replacing Components of the Manifold
Thermo Scientific
TSQ 9000 Hardware Manual
259
Front Manifold Plate
To replace the front manifold plate:
1. Shut down your system by following the steps in
“Shutting Down the TSQ 9000 System”
.
2. Remove the vacuum interlock knob and use a T10 Torxhead screwdriver to remove the
five screws around the knob. Replace the knob and screws, if necessary.
Note
See the
ISQ and TSQ GC-MS Spare Parts Guide
for information about ordering this
component.
Note
You do not have to remove the ion source cartridge or the column.
Note
As you remove the screws around the vacuum interlock knob, watch for a small
spring and safety interlock (location shown below) that will fall out as your remove the
knob. Be sure to catch them and set them aside. Replace the spring and safety interlock, if
necessary.
Figure 273.
Removing the Vacuum Interlock Knob
Screws
Spring and safety interlock
are behind this screw