3
Advanced Troubleshooting
Replacing Components of the Vacuum Interlock
238
TSQ 9000 Hardware Manual
Thermo Scientific
Replacing Components of the Vacuum Interlock
If the components of your vacuum interlock get damaged, you can use the illustration below
to visually locate a component and then follow the process to replace it.
To replace the components of the vacuum interlock
1. Shut down your system by following the steps in
“Shutting Down the TSQ 9000 System”
.
Note
See the
ISQ and TSQ GC-MS Spare Parts Guide
for information about ordering
these components.
Figure 243.
Replaceable Components of the Vacuum Interlock
Vacuum Interlock Knob
Inner Ball Seal
Sealing Ball
Axle Seal
Axle Bearing
Valve Axle
Clip
O-Ring
Spring
(Metal valve body is
shown for illustration
purposes only. It is not
available for purchase.)
O-Ring
Outer Ball Seal