Advanced Troubleshooting
Replacing Components of the Ion Source
Thermo Scientific
TSQ 9000 Hardware Manual
225
Figure 221.
Removing the Ion Source Cartridge
13. Let the ion source cartridge cool down before removing the components from the source
holder.
14. Set the ion source cartridge and holder on a clean surface
15. Replace the ion source cartridge or any individual component.
16. Follow the steps in
Reassembling the Extractabrite Ion Source Cartridge
.
Reinserting the Extractabrite Ion Source Cartridge into a TSQ 9000 System without
a Vacuum Interlock
CAUTION - BURN HAZARD
The ion source cartridge may be hot.
AVERTISSEMENT RISQUE DE BRÛLURE
:. la cartouche de la source d'ions peut être
chaude.
IMPORTANT
When inserting a cold ion source cartridge such as after cleaning, the ion
source and lens stack will expand as the source cartridge heats, often pushing the ion
volume and lenses away from the rear of the instrument where they are firmly held by the
RF Lens spring contacts. To avoid intermittent electrical contacts to the lenses, you should
insert the ion source cartridge, wait 30 minutes for it to get to temperature, then remove
and reinsert it.