3
Advanced Troubleshooting
Replacing Components of the Ion Source
Thermo Scientific
TSQ 9000 Hardware Manual
181
Thumbscrew-Repeller Spacers
To replace the thumbscrew-repeller spacers
1. Shut down your system by following the steps in
“Shutting Down the TSQ 9000 System”
.
2. Use a T20 Torxhead screwdriver to remove the four screws around the manifold door.
3. Open the manifold door.
Note
See the
ISQ and TSQ GC-MS Spare Parts Guide
for information about ordering this
component.
Note
Remove the screws completely so that they do not scratch the manifold when you
close the door.
Figure 161.
Opening the Manifold Door
Screws