28 | Touchscreen Controller Display Functioning
Medical Devices -Touch Controller
The second is called USER: It is enabled by default to do
everything using password 1234.
New users can be created by gently pressing “NEW”
and typing in the name (for example JOHN). After
confirming by pressing,
a specific password will
be requested for this user.
Double click the name of the created user (if selected
correctly, it will turn blue) and enable or disable the
functions required for this user (setpoint and limits
changing, switching on/off the equipment, the door
opening, the date/hour changing, etc.).
ScreenSaver
User can enable screensaver entering software panel
setup. Screensaver will automatically activate, if the
screen is not in use for a certain period of time.
The screensaver page will show the chamber
temperature against a black background easily readable
from the distance. The screensaver page will disappear
at the touching of the screen.
Test Function
Test program allows to check operation of some
security signaling, such as acoustic alarm for event of
high temperature, acoustic alarm for event of low
temperature and remote alarm test for acoustic alarm
and dry contacts.
During these tests, device operation keeps taking place
regularly and temperature inside the chamber remains
within limits.