Thermo Scientific
TruDefender FT and TruDefender FTi User Guide
11
2
Scanning
Best Practices for Scanning
For this situation
Do these things
Puddled liquid
At least 1/4 inch (6 mm) deep
Immerse the sampling tip in the
puddle. The instrument may touch the
underlying surface.
Non-volatile liquid in a
container
Stand the instrument on its bottom.
Place a drop of liquid on the sampling
tip.
Volatile liquid
See
Thinly spread film or
droplets of liquid
Saturate a swab with the liquid.
Press the swab
firmly
on the center of
the sampling tip.
Thinly spread paste
Collect some of the paste on a swab.
Press the swab
firmly
on the center of
the sampling tip.