Chapter 12
Replace the Collecting Electrode
Maintaining an Electron Capture Detector
Maintenance and Troubleshooting Manual
229
OPERATING SEQUENCE
Replace the Collecting Electrode
Materials needed
•
Screwdriver
•
2.5 mm Allen wrench
•
Tweezers
•
Collecting Electrode (P/N 277 003 50)
WARNING! Any maintenance or repair operations involving the radioactive source must be
performed ONLY by qualified personnel duly authorized to handle radioactive
material. Please contact your local Thermo Fisher Scientific Technical Service office.
WARNING! When handling organics solvents you must take precautions to avoid health
hazards. Refer to Appendix A,
for more information.
WARNING! This operation must be carried out at low temperature to avoid burns. Therefore,
before beginning the sequence, the detector must be cooled to room temperature.
PRECAUTIONS