Chapter 7
Maintaining a Packed Column Injector
Packed Column Injector Maintenance
124
Maintenance and Troubleshooting Manual
Packed Column Injector Maintenance
The Packed Column injector and its interface module will normally be serviced by
Thermo Fisher Scientific authorized technical personnel. In order to keep the
injector operating at peak performance, you should periodically replace the
septum.
When replacing the septum
The septum must be replaced at least every 200 injections, or every time a
problem occurs related to a septum damage or wear. Refer to Chapter 19,
, for more information. Instructions are reported in the