1
Installation
Installing the Column the First Time
Thermo Scientific
TRACE 1300 and TRACE 1310 Hardware Manual
57
Figure 66.
Column Rack
To install a column rack
1. Open the front door of the GC.
2. Slightly press the sides of the column rack and insert the two hooks into the
corresponding button-holes on the oven heater baffle.
Installing the Column the First Time
This section contains instruction for installing the capillary column into the GC oven, for
connecting the column ends to the injector and the detector and for carrying out the
column
conditioning
, the
leak test
and the
column evaluation
.
To install a new column the first time
1. Install the column into the GC oven.
a.
Open the front door of the GC.
b. Place the column on the two arms of the rack. See
.
Column Rack
Hooks
Button-holes
Oven Heater
Baffle
Arms
CAUTION
The injector and detector modules are shipped with a plug.
Before installing the column, remove the plug from the injector and detector modules.
Unscrew the column retaining nut from the bottom of the injector and the detector
modules, then remove the blind ferrule.