Adding Modules
Adding an Aux Temperature/Cryo Module
Thermo Scientific
TRACE 1300 and TRACE 1310 Hardware Manual
447
Figure 465.
Aux Temperature/Cryo Module Installed into the GC
8. Connect the Aux Temperature/Cryo module.
a.
Using the cable provided, connect the 15-pin female connector marked
GC Bus
on
the module to a
Bus
interface on the back of the GC.
–
If a DSQ II, ITQ, or TSQ Quantum mass spectrometer is coupled with the GC,
connect the heater cable coming from the transfer line to the connector marked
Heater 1
or
Heater 2
.
–
If the cryo option for the Oven is present, connect the 2-pin connector marked
Cryo Valves - Oven
to the cryo solenoid valve using the cable provided.
–
If the cryo option for the front/back PTV/PTVBKF injector is present, connect
the 2-pin connector marked
Cryo Valves - Front Inlet
or
Cryo Valves-Back
Inlet
to the cryo solenoid valve using the cable provided.
–
If the on/off activation of external solenoid valves (up to eight) is required,
connect a 2-pin connector marked
Event 1÷8
to each external solenoid valve.
–
Plug the power cable to the AC Input connector on the front of the module and
to the wall outlet. The LED marked
On
lights after the GC is powered on.
Tip
To avoid the contact with the hot air from the vents, it is suggested to gather the
electrical cables into cables holder.
Aux Temperature/Cryo Module
Cables Holder
ATTENTION
An extension cable transfer line is needed when connecting an ITQ mass
spectrometer to the Aux Temperature/Cryo module.