Injectors Advanced Maintenance
Cleaning the SSL Injector Body
Thermo Scientific
TRACE 1300 and TRACE 1310 Hardware Manual
235
Figure 219.
SSL Injector Components
9. Remove the body head o-rings. See
.
a.
Using tweezers, remove both the internal and external body head o-rings.
b. Place and keep all the removed components on a clean surface.
10. Remove the injector body. See
.
a.
Using a T20 Torxhead screwdriver, undo the two injector body fixing screws, and
extract the injector body from its housing.
Septum Cap
Ring Nut
Septum Holder/Liner Cap with Septum
Liner
Liner Seal
Bottom Fittings
Retaining Nut
Washer
Base Seal
Ferrule
Terminal Fitting for Capillary Column
Capillary Column Retaining Nut