5.5.2
Flow Connections Overview
The picture illustrates the liquid flow path through the system:
Figure 22: Flow connections in the Vanquish system (example)
The Vanquish system has been designed for optimized and simple leak
liquid and waste handling:
•
Leak liquid from the solvent rack, detector, autosampler, and pump
flows through pipes from the modules to the system base drain port.
•
Seal wash and needle wash liquids from the pump and autosampler,
as well as condensate from the autosampler, are also guided to the
system base drain port.
•
Leak liquid from the column compartment flows to the column
compartment drain port.