5
•
Installation
UHPLC System
Operating Manual
Page 57
Follow these steps
1. On the autosampler, pump, and column compartment, connect the
"B" connector of a USB cable to the
USB
port.
2. Connect the "A" connector of the USB cable from the pump to the
USB
hub on the UV/VIS detector. Connect the "A" connector of the
USB cable from the other modules to the
USB
hub on the UV/VIS
detector or pump.
3. Connect the "B" connector of a USB cable to the
USB
port on the
UV/VIS detector, and connect the "A" connector of the USB cable to
a USB port (2.0 or higher) on the computer.
In order to control the modules that use a USB hub on another module,
the module that provides the hub must be turned on.
NOTICE
•
Never use defective communication cables. If you suspect that a
cable is defective, replace the cable.
•
To ensure trouble-free operation, use only the cables provided by
Thermo Fisher Scientific for connecting the system.
5.4.2.3
System Interlink Connections
The system interlink connections require a special system interlink cable,
(Cat. 6 cable).
Figure 13: System interlink cable
Summary of Contents for thermoscientific Vanquish UHPLC System
Page 4: ...Contacting Us Page 4 UHPLC System Operating Manual ...
Page 28: ...2 Safety Page 28 UHPLC System Operating Manual ...
Page 34: ...3 System Overview Page 34 UHPLC System Operating Manual ...
Page 38: ...4 Unpacking Page 38 UHPLC System Operating Manual ...
Page 82: ...5 Installation Page 82 UHPLC System Operating Manual ...
Page 100: ...6 Operation Page 100 UHPLC System Operating Manual ...
Page 112: ...7 Maintenance and Service Page 112 UHPLC System Operating Manual ...
Page 128: ...8 Troubleshooting Page 128 UHPLC System Operating Manual ...
Page 132: ...9 Specifications Page 132 UHPLC System Operating Manual ...
Page 144: ...11 Appendix Page 144 UHPLC System Operating Manual ...