5
•
Installation
Page 66
Pumps C, Pumps F ( VC-Pxx, VF-Pxx)
Operating Manual
1. Assemble the solvent line filter, wearing appropriate clean gloves:
a) Place the frit in the filter holder (bottom part).
b) Make sure that the frit is in a level position.
c) Screw the filter top to the filter bottom.
2. On the pump inlet, remove the plug that protects the inlet during
shipment, if applicable.
TIP
For easy access to the lower degas chambers, remove the seal wash
detector assembly (funnel and detector). Grasp and pull the assembly
upward by the funnel. The detector including the tubing and cable can
remain connected to the funnel.
Figure 17: Removing the seal wash detector assembly
No.
Description
1
Funnel, seal wash detector
2
Seal wash detector
3
Detector cable
4
Seal wash line
Follow these steps
Connecting the solvent lines comprises the following steps:
1. Routing the solvent lines to the solvent rack
2. Connecting the solvent lines to the solvent reservoirs
3.
Optional:
Installing shut-off valves in the solvent lines