9
•
Pump-Specific Information
Pumps C, Pumps F ( VC-Pxx, VF-Pxx)
Operating Manual
Page 223
9.2.4
Setting the Solvent Composition
This section describes how you set the solvent composition for the
quaternary pump (solvents A to D). The steps and examples apply
equally to the ternary pumps of the dual pump (solvents AR to CR and
AL to CL).
Follow these steps
1. In the Instrument Configuration Manager, in the dialog box for the
pump, check and the change the following if required:
¨
Number of the solvents to be used with the pump
¨
Standard solvent names
These solvent names appear in the Chromeleon user interface.
2. In the Chromeleon Client, determine the percentage of solvent that
you want to deliver.
Set the percentage for solvents B, C, and D (
%B, %C, %D
). Solvent A
(
%A
) automatically delivers the remaining volume.
Examples
•
To deliver 100% of solvent
%A
:
Set
%B, %C
, and
%D
to 0%.
•
To deliver 100% of solvent
%B
(or
%C
or
%D
):
Set
%B
(or
%C
or
%D
) to 100%. The remaining solvents are
automatically set to 0%.
•
To deliver 20% of solvent
%A
, 40% of solvent
%B
, 30% of solvent
%C
,
and 10% of solvent
%D
:
Set
%B
to 40%, set %
C
to 30%, and set
%D
to 10%.
%A
is
automatically set to 20%.
Summary of Contents for thermoscientific Vanquish Pumps C
Page 4: ...Contacting Us Page 4 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 30: ...2 Safety Page 30 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 40: ...3 Device Overview Page 40 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 196: ...7 Maintenance and Service Page 196 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 214: ...8 Troubleshooting Page 214 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 290: ...13 Appendix Page 290 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...