Variable Wavelength Detectors (VC-D40, VF-D40)
Operating Manual
6.7
Important Operating Parameters
If a parameter listed below is not available in the Chromeleon software,
consider updating the firmware and Chromeleon version.
TIP
The Instrument Method Wizard provides different parameter view
modes. Depending on the parameters to be set, you can select the
desired view mode (for example,
Easy
or
Advanced
).
For more information, refer to
Chromeleon Help and documents
.
Parameters
Description
Data collection rate
Sets the number of data points per second (Hz) that the
Chromeleon software collects from the detector and
stores as raw data. The maximum data collection rate
depends on the detector type and the software version.
For more information on how to set the data collection
rate, see
.
Lamphouse
temperature
The temperature signal channel (
Lamphouse
Temperature
) that is available in the Properties dialog
box of the detector, can be used for troubleshooting
purposes.
The lamps of the detector are very sensitive to changes in
temperature. In case of baseline drift problems, the lamp-
house temperature signal can help to find the reason for
the problem.
Select the signal channel to record the temperature.
Chromeleon generates the appropriate channel for
recording the temperature of the detector lamp house.
Leak detection
Leak detection is enabled as a standard when the
detector is shipped (
Leak Sensor
Mode = Enabled
). This is
the preferred setting.
Peak width
The peak width is a supporting parameter for the data
collection rate setting and the response time.
For details on the peak width and selecting it individually,
see
Response Time and Peak Width (
.
Response time
The response time determines how quickly the detector
responds to a change in signal.
For details on the response time and selecting it
individually, see
Summary of Contents for Thermo Scientific Vanquish VC-D40
Page 4: ...Contacting Us Page 4 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Page 10: ...Contents Page 10 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Page 98: ...6 Operation Page 98 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Page 148: ...8 Troubleshooting Page 148 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Page 154: ...9 Specifications Page 154 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Page 168: ...11 Appendix Page 168 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Page 173: ...Index Variable Wavelength Detectors VC D40 VF D40 Operating Manual Page 173 ...