![Thermo Scientific Thermo Scientific Vanquish VC-D40 Operating Manual Download Page 79](http://html.mh-extra.com/html/thermo-scientific/thermo-scientific-vanquish-vc-d40/thermo-scientific-vanquish-vc-d40_operating-manual_1105377079.webp)
6
•
Operation
Variable Wavelength Detectors (VC-D40, VF-D40)
Operating Manual
Page 79
MUTE ALARM
Beeping alerts you when the device detects a problem, for example,
leakage. To turn off the beep for the current alarm, press this button.
Eliminate the source for the alarm within 10 minutes. Otherwise,
beeping starts again. If the device detects a different problem, beeping
alerts you again immediately.
UV
The
UV
button allows you to turn the UV lamp on and off directly from
the detector. To turn on the UV lamp, press the
UV
button.
The LED next to the button indicates the lamp status:
LED
Description
Off (dark)
The UV lamp is turned off.
Green, flashing
The UV lamp ignites.
Green
The UV lamp is turned on.
VIS
The
VIS
button allows you to turn the VIS lamp on and off directly from
the detector. To turn on the VIS lamp, press the
VIS
button.
The LED next to the button indicates the lamp status:
LED
Description
Off (dark)
The VIS lamp is turned off.
Green, flashing
The VIS lamp ignites.
Green
The VIS lamp is turned on.
When the device is connected in the Chromeleon software
The button functionality is as follows when the device is connected in
the Chromeleon software:
•
No injection or sequence or baseline monitoring is running:
All functions are available from the keypad.
•
An injection or sequence or baseline monitoring is running:
The
MUTE ALARM
function remains available from the keypad,
allowing you to turn off the beep for the current alarm.
6.3.2
Status Indicators
The status LED bar on the front side of the device and the
STATUS
LED
on the inside keypad provide information about the device status.
Summary of Contents for Thermo Scientific Vanquish VC-D40
Page 4: ...Contacting Us Page 4 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Page 10: ...Contents Page 10 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Page 98: ...6 Operation Page 98 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Page 148: ...8 Troubleshooting Page 148 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Page 154: ...9 Specifications Page 154 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Page 168: ...11 Appendix Page 168 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Page 173: ...Index Variable Wavelength Detectors VC D40 VF D40 Operating Manual Page 173 ...