7.13 Replacing the Slide-In Module
7.13.1 Removing the Slide-In Module
CAUTION—Heavy Load, Bulky Device
The device is too heavy or bulky for one person alone to handle safely.
To avoid personal injury or damage to the device, observe the following
guidelines:
•
Physical handling of the device, including lifting or moving, requires
a team effort of two persons.
•
A team effort is in particular required when lifting the device into
the system stack or when removing it.
Tools required
Screwdriver, Torx T20
Preparations
1. Prepare the device for transport. See
.
Follow these steps
1. Loosen the four captive screws on the front left and front right of
the device.
Figure 28: Captive screws on the slide-in module
2. Push all tubing and capillaries, which are present in the tubing chase
of the Vanquish system modules, into the tubing chase. Otherwise,
you will not be able to remove the slide-in module properly from the
enclosure in the next step.
3. Grasp the slide-in module by the leak tray, or by the lamp house
cover and the partition panel, and pull the module out of the
enclosure by approximately 10 cm.
Summary of Contents for Thermo Scientific Vanquish VC-D40
Page 4: ...Contacting Us Page 4 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Page 10: ...Contents Page 10 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Page 98: ...6 Operation Page 98 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Page 148: ...8 Troubleshooting Page 148 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Page 154: ...9 Specifications Page 154 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Page 168: ...11 Appendix Page 168 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Page 173: ...Index Variable Wavelength Detectors VC D40 VF D40 Operating Manual Page 173 ...