2
Installation
Connecting Remote Outputs
Thermo Scientific
Surveyor PDA Plus Detector Hardware Manual
21
Figure 17.
Configuration page of the PDA Direct Control dialog box in Xcalibur
Changing the Polarity of the Remote Outputs from ChromQuest
To change the polarity of the remote outputs
1. Install the ChromQuest data system, if you have not already done so.
2. Add the Surveyor PDA Plus Detector to the instrument configuration (see
the Surveyor PDA Plus Detector in ChromQuest”
).
3. Open the online Instrument window.
4. Choose
Control > Instrument Status
.
5. Click the
Surveyor PDA
tab.
6. Click the appropriate button in the Output States area on the PDA detector page.