4
Start-up
Gas connection
Thermo Scientific
S
TERI-CYCLE
i160
/
i250
4-11
4. Secure the gas pressure hose to the sleeve of the sterile filter using the hose clamp.
Gas connection without gas monitoring system
Figure 4-11.
Gas connection without gas monitoring system
Note
Pressure compensation opening:
To ensure permanent pressure compensation, the pressure compensation opening must not be
connected to an exhaust air system. The pipe of the pressure compensation opening must not be
extended or redirected.