12
Annex: Data communication
Overview of general parameters (addresses 0xxx)
Thermo Scientific
S
TERI-CYCLE
i160
/
i250
12-5
Description of the response elements:
aaaa
parameter address,
bb
amount of payload (always 02)
cc
Checksum: Inverted XOR of all bytes without
checksum and <CR>
XX = 2 bytes error message (see table below)
Example of an unknown command:
Query:
?:0005:00::cc<CR>
Response
!:0005:02:?1:cc<CR>
Significance of the two bytes in the error message:
Overview of general parameters (addresses 0xxx)
General parameters are system values such as date, time, and motherboard version number.
Reading general parameters
Overview of incubator parameters (addresses 2xxx)
Incubator parameters are divided into:
• the basic parameters of the three control loops temperature, CO
2
and O
2
,
• the internal function parameters of operating functions and data logging.
Failure message
Description
?0
Error in telegram structure or check sum
?1
Unknown error or unknown parameter
?2
Internal memory error
?3
Data error (value not within set limits)
Address
Description
Remarks
0001
Mainboard version number
8 places
0010
Date and time display
[hours:minutes:seconds];[day:mon
th:year]
17 bytes / decimal value
with the format xx:xx:xx;xx:xx:xx
0011
Date
[day:month:year]
8 bytes / decimal value
with the format xx:xx:xx
0012
Time
[hours:minutes:seconds]
8 bytes / decimal value
with the format xx:xx:xx
Summary of Contents for STERI-CYCLE i160
Page 8: ...Table of Contents 6 STERI CYCLE i160 i250 Thermo Scientific ...
Page 12: ...Figures 4 STERI CYCLE i160 i250 Thermo Scientific ...
Page 76: ...5 Operation Starting operation 5 6 STERI CYCLE i160 i250 Thermo Scientific ...
Page 154: ...9 Maintenance Replacing the Door Seal 9 12 STERI CYCLE i160 i250 Thermo Scientific ...
Page 166: ...11 Specifications STERI CYCLE i250 11 10 STERI CYCLE i160 i250 Thermo Scientific ...
Page 196: ...13 Device log 13 2 STERI CYCLE i160 i250 Thermo Scientific ...