Handling and control
User configuration
6-36
S
TERI-CYCLE
i160
/
i250
Thermo Scientific
Figure 6-40.
Setting the water level sensor
1. To toggle between two states:
• Press the + key.
or
• Press the - key.
2. To accept and save the change:
• Press the
Enter
key.
• The system returns to the
Options
menu.
Switching the audible alarm on/off:
If the device-integral monitoring system detects an error:
• an audible alarm sounds in addition to
• the visual error message and the switching of the alarm relay.
The audible alarm can be permanently disabled.
Switching the audible alarm on/off:
1. Press the
Menu
key.
2. Select the menu command
Options
.
• The selection dialog shown in
3. Select the menu command Buzzer.