Chapter 5
Configuration
6. Continue this selection process for up to six validation steps.
V a l i d a t i o n
S e t u p
S t e p
6
S e l e c t i o n
∨
∧
E n t
E x i t
Figure 5-52.
7. Enter the number of minutes (0-60) that the SOLA II Flare is to remain
in this particular validation step.
V a l i d a t i o n
S e t u p
S t e p
6
T i m e
∨
∧
E n t
E x i t
Figure 5-53.
4–20 mA Range
Setup
Each calibration is configured to use a unique range for the 4–20 mA output.
The low end of the range (4 mA) always corresponds to zero sulfur
concentration, and the upper end of the range (20 mA) is configurable for each
calibration.
1. Enter the 20 mA value for the Calibration A Range.
4
t o
2 0
m A
R a n g e s
C a l i b r a t i o n
A
R a n g e
∨
∧
E n t
E x i t
Figure 5-54.
Thermo Scientific
SOLA II Flare User
Manual
5-20