SimpliAmp

 Thermal Cycler PCR System Installation and

Operation

Catalog Number 

A24811

Pub. No. 

A24827   

Rev. 

D

Note: 

For safety and biohazard guidelines, see the “Safety”

appendix in the SimpliAmp

 Thermal Cycler User Guide (Pub. No.

MAN0009889). Read the Safety Data Sheets (SDSs) and follow the

handling instructions. Wear appropriate protective eyewear,

clothing, and gloves.

Product description

This document summarizes procedures for installing and using

the SimpliAmp

 Thermal Cycler, with "better than gradient

"

VeriFlex

 technology. For detailed instructions, see the SimpliAmp

Thermal Cycler User Guide (Pub. No. MAN0009889).

Figure 1    SimpliAmp

 Thermal Cycler

Required materials not provided

• Can of compressed air (For use in cleaning wells of sample

block).

• (Optional) Electrical protective devices.

Note: 

The use of one or more of the following electrical

protective devices is recommended.

·

Power line regulator (100–240 V)

·

Surge protector/line conditioner (10-kVA)

·

Uninterruptible power supply (1.5-kVA)

Set up the SimpliAmp

 Thermal Cycler

IMPORTANT! 

Save the packing materials and box in case you

need to ship the instrument in for service.

1.

Open the shipping crate to unpack the instrument. You

should receive one box containing the thermal cycler and the

accessories.

2.

Remove the packing material, then inspect the instrument for

shipping damage.

3.

Use compressed air to clear out each well of the sample block

to remove particles that may have collected inside during

shipping.

4.

Place the instrument at an installation site that meets the

spatial and weight requirements for the thermal cycler (see 

“Technical specifications“ on page 6).

5.

Connect the instrument to a power outlet.

a.

Connect the power cord to the instrument.

1

Power cord port

b.

Install any desired electrical protective devices.

c.

Connect the power cord to a wall plug.

6.

Press the power button, then wait for the instrument to start

up. Proceed with the installation after the touchscreen

displays the Home screen, indicating that the instrument is

active.
When you power on the instrument, the instrument will take

under a minute to start up.

QUICK REFERENCE

For Research Use Only. Not for use in diagnostic procedures.

Summary of Contents for SimpliAmp

Page 1: ...Set up the SimpliAmp Thermal Cycler IMPORTANT Save the packing materials and box in case you need to ship the instrument in for service 1 Open the shipping crate to unpack the instrument You should re...

Page 2: ...create an account Link the PCR instrument to Thermo Fisher Connect administrator only The procedure to link the Administrator profile to Thermo Fisher Connect only needs to be performed the first tim...

Page 3: ...Connect by Thermo Fisher Scientific 2 For Android devices download the application from Google Play by searching for Instrument Connect by Thermo Fisher Scientific 3 Log in to the instrument connect...

Page 4: ...bes are placed in the sample block without support they can be crushed or deformed when the lid of the thermal cycler is closed over the block Using the tray and retainer set allows the pressure to be...

Page 5: ...nward pressure on the sealing tool in all steps to form a complete seal on top of the tubes 1 Align and place the cap strips on the tubes 2 Seal the cap strips using the rocking capping tool a Slip yo...

Page 6: ...V 50 60 Hz Max 600 W VeriFlex blocks 3 VeriFlex zones Supported 10 C 5 C zone to zone Settable 3 20 C 10 C zone to zone Ambient humidity 15 80 Relative Humidity non condensing acceptable range Ambient...

Page 7: ...Optical 384 Well Clear Reaction Plates with Barcode 4483285 4483273 Strip tubes and caps MicroAmp Fast 8 Tube Strip 0 1 mL 4358293 MicroAmp Optical 8 Tube Strip with Attached Optical Caps 0 2 mL A305...

Page 8: ...without notice DISCLAIMER TO THE EXTENT ALLOWED BY LAW THERMO FISHER SCIENTIFIC INC AND OR ITS AFFILIATE S WILL NOT BE LIABLE FOR SPECIAL INCIDENTAL INDIRECT PUNITIVE MULTIPLE OR CONSEQUENTIAL DAMAGES...

Reviews: