User Mode
The basic operation of the SAM12 is as follows:
•
Upon power up, system diagnostics are run – see
.
•
Upon successful system diagnostics, the SAM12
establishes a new Background – see
.
•
The system enters the Background Monitoring Mode –
see
Normal Background Monitoring(page 5-49)
.
•
A user places articles to be monitored inside the vault
and closes the door – see
.
•
Monitoring starts as a result of the Start button being
pressed – see
Commence Monitoring (page 5-53)
.
•
Monitoring continues for a period determined by the
required confidence levels and the current background.
•
Monitoring may be interrupted due to the result of a
QuickScan assessment (see
), or if a
significant change in background conditions is detected
during the measurement – see
during Measurement (page 5-54)
.
•
At the end of the Monitoring period, activity is
calculated and compared against a set of alarm
thresholds. A simple Go/No Go indication is given
along with details of the measured activity. If the
activity is above the alarm threshold, NBR algorithms
are used to attempt to determine the proportion of
activity that is due to NORM – see
(page 5-55)
.
•
The user removes the article from the SAM12 - see
.
Summary of Contents for SAM12
Page 1: ...MANUAL SAM12 SMALL ARTICLES MONITOR Thermo Fisher Scientific SAM12 21st September 2007...
Page 2: ......
Page 4: ...This manual was produced using ComponentOne Doc To Help...
Page 10: ......
Page 18: ......
Page 20: ......
Page 34: ......
Page 74: ...Operating Instructions Issue 1 0 5 34 SAM12 User Manual Thermo Fisher Scientific...
Page 130: ......
Page 134: ......
Page 144: ......
Page 168: ......