Operating Instructions
Issue: 1.0
5-24 SAM12 User Manual
Thermo Fisher Scientific
To check the current
Calibration
, select a Calibration
Source and Calibration Accuracy.
Then press Cal Check button to
display the Ready to Start
window.
Remove all items from the
SAM12 and press the Start
button.
Note that the Abort button can be
pressed at any time during the
process to return to the previous
window.
When the Start button is pressed,
the process is similar to that
described above in
.
Description of Error messages:
Error Message
Resolution
Unable to calculate a
valid monitoring time
1. There is no source available
2. The source does not have enough activity
3. If a cobalt calibration has been activated and the
source is not cobalt60
This is not the default
Calibration Source
This is a warning message only
Select Calibration|Calibration
Selection to view current
Calibration and create new
customised Calibration
Note that the Active Calibration with its Gross Sum Efficiency
% is displayed in the top box.
To change the Active Calibration
, use the direction arrows to
select a Calibration. Once selected, press Make Active button.
The selected Calibration and its Gross Sum Efficiency will be
displayed in the top box.
Calibration Selection
Summary of Contents for SAM12
Page 1: ...MANUAL SAM12 SMALL ARTICLES MONITOR Thermo Fisher Scientific SAM12 21st September 2007...
Page 2: ......
Page 4: ...This manual was produced using ComponentOne Doc To Help...
Page 10: ......
Page 18: ......
Page 20: ......
Page 34: ......
Page 74: ...Operating Instructions Issue 1 0 5 34 SAM12 User Manual Thermo Fisher Scientific...
Page 130: ......
Page 134: ......
Page 144: ......
Page 168: ......