Maintenance and Care
Cleaning
Thermo Scientific
Sorvall RC 3BP Plus
7-3
Clean centrifuge and accessories as follows:
1. Open the centrifuge.
2. Turn off the centrifuge.
3. Pull out the power supply plug.
4. Prepare the rotor according to the rotor manual.
5. Remove the centrifuge tubes and adaptors.
6. Use a neutral cleaning agent with a pH value between 6 and 8 for cleaning.
7. Dry all of the rotors and accessories after cleaning with a cloth or in a warm air cabinet at a
maximum temperature of 50° C.
8. Clean the housing of the centrifuge as needed.
•
After cleaning, treat the entire surface of aluminum parts with corrosion protection oil
(70009824). Also treat the cavities with oil.
•
Tread the bolt of the swing out rotor with bolt grease (75003786).
Defrost rotor chamber
After some applications there might be ice in the rotor chamber.
1. Install a rotor and close the centrifuge.
2. Set run parameters to SPEED = 2500 rpm, TIME = 5:00 (longer for heavy frost buildup),
TEMP = 30.
3. Press the
key.
Clean the rotor chamber as described above.
CAUTION
Observe that no liquids, particularly no organic solvents, get into the openings
at the motor shaft outlet, as they could leach out the grease of the bearings.
CAUTION
Lift the rotor with two persons due to its weight.
CAUTION
When cleaning, do not allow liquids, especially organic solvents, to get on the
drive shaft, the bearings, or the locks.
Organic solvents break down the grease in the motor bearing. The drive shaft could freeze
up.