Thermo Scientific PCTestr 35 Instruction Manual Download Page 1

INSTRUCTION MANUAL

 

MULTI-PARAMETER TESTR 35 SERIES

 

pH / Conductivity / TDS / Salinity / Temperature 

  

 

 

  

 

 

Part of Thermo Fisher Scientific     68X441601 Rev. 1 March 2010 

 

Thank you for selecting our Multi-Parameter Testr. This manual serves the following (3) models: 

• PCTestr 35 (Eutech PCTEST35-01X441504 / Oakton 35425-00) 

 pH / Conductivity / Temperature 
 

• PTTestr 35 (Eutech PCTEST35-01X441505 / Oakton 35425-05) 

pH / Total Dissolved Solids / Temperature 

 

• PCSTestr 35 (Eutech PCSTEST35-01X441506 / Oakton 35425-10) 

pH / Conductivity / Total Dissolved Solids / Salinity / Temperature 

 

Getting Started:

  

Your instrument has been factory calibrated and usually works well out of the box. However, after 

extended periods of non-use, it’s best to remove the sensor cap and soak the 
sensor (pictured here) in warm tap water or pH buffer for 10 minutes or so. A 
brief rinse with deionized (DI) water is OK, but avoid soaking or storing in 
deionized water as this will shorten the pH electrode life.  Prior to taking 
measurements, periodic calibration with certified standards is recommended for 

best accuracy. 
 
Your Testr begins in the measuring mode that was previously used. Just prior to measurement or 
when switching modes, you will see the setting associated with each parameter i.e.) pH (buffer 
group selected), Conductivity (Auto), TDS (factor), Salinity (unit of measure).  
 

Setup:

  

Your Testr allows customization of various settings. To access the setup mode:  

1.

 

With the Testr off, keep the 

 pressed down while you press and release 

.  

 (Setup) will appear, then as you release 

 (Parameter) will appear.  

2.

 

Press 

 or  

 

to choose 

 (Parameter Setup) or 

 (System Setup) 

menu.  

3.

 

Press 

 to enter the selected setup menu.

 

 

 

*IMPORTANT*

 It is necessary to save your Parameter and System 

changes in order for them to take effect. 

 

1.

 

When you are finished making your desired changes, press both 

and 

at the same 

time and keep them pressed until you see “SA” (Save) on the display.  

2.

 

With the primary display “SA” and secondary display “YES”, press 

 to save the 

changes. The instrument will resume measurement mode with new setting(s).

 

 

 

Note: If auto-shut off is used, changes will be automatically saved 8.5 minutes after the last

 

change was made.

 

Parameter Setup: 

Select 

 to make changes relating to the 

parameters – pH, Conductivity TDS, Salinity. Note: only the PCS Testr will have all of 
these options. See below for menus available from each parameter.  To Navigate the 

menus: 

 

Press 

 to select or confirm the displayed option.  

 

Press 

 or  

 to scroll thru options or change values. 

 

pH Options:  

 

USA or NIST Buffer Group for calibration buffer option. 

 

5-pt calibration (all points) or 3-pt calibration (middle three points only). 

 

Salinity (SALt) Option (PCS Testr only) 

 

Choose PPt (parts per thousand) or Per (percentage %) as unit of measure. 

 

Total Dissolved Solids (tDS) Option (PT and PCS Tester only) 

 

FACt factor the instrument uses to convert from conductivity to TDS value. 

Adjustable from 0.40 to 1.00 (default factor is 0.71). 
 

Conductivity Options (PC and PCS Tester only) 

 

A.Cal (Automatic Calibration) Choose YES or NO (manual). 

TIP:

  The PC Testr 35 and PCS Testr 35 are capable of automatic or manual 

conductivity calibration. In automatic calibration mode, the meter will automatically 
choose one of (3) conductivity calibration standards depending on the ranges listed 

below. If you will only use 84 

µ

S, 1413 

µ

S, or 12.88 mS calibration standards, 

automatic calibration is a time saving option. If you intend to calibrate with one or more 
standards that are not listed below, choose “NO” which will disable auto calibration and 

allow you to enter your desired value manually. 

Conductivity Range Automatic Calibration Value  Available with

0.0 – 200.0 

µ

S 84 

µ

S PCS 

only 

201 – 2000 

µ

S 1413 

µ

PC or PCS 

2.01 – 20.00 mS 

12.88 mS 

PC or PCS 

 

SPC (Single-Point Calibration) Choose YES or NO (multi-point calibration). 

TIP:

 

The PC Testr 35 and PCS Testr 35 are capable of single or multi-point conductivity 

calibration. Use Single-Point Calibration to apply a single calibration value across all 
ranges. Use Multi-Point Calibration for individual calibration in each range. This will 

restrict an individual calibration so that it is applied to one range only. When using 
multi-point calibration, perform a calibration in each range that you expect to use for 

best results. 
 

System Setup: 

Select 

 to make changes relating to the system.  See below for available menus.  

Note: other than changing Temperature units, it is advised to keep the factory default 

settings for best results. To Navigate the System menus: 

 

Press 

 to select or confirm the displayed option.  

 

 

Press 

 or  

 to scroll thru menu options or change values. 

 

 

Unit rSt (Instrument reset) 

 

PH (pH) or EC (electrical conductivity / TDS / Salinity) 

 

CAL (calibration reset) or FCt (Reset to factory default settings) 

 

Set A.Off (Automatic shut off after 8.5 minutes) Choose YES or NO. 

 

Set t.C (Temperature Coefficient) 0.0-10.0% (2.1% is default) 

 

Set AtC (Auto Temperature Compensation) Choose YES or NO (25°C is used). 

 

Set °C °F (select temperature units) Choose °Celsius or °Fahrenheit. 

 
 

Temperature Calibration:

  

The factory temperature calibration should last for the life of the original sensor since it 

doesn’t normally drift. Temperature calibration is always recommended upon sensor 
replacement. It may also be desirable to adjust the temperature to match a certified 

accurate thermometer or another Testr. The temperature value is common to all 
parameters so only one calibration is needed. To perform temperature calibration: 

1.

 

Press 

 to turn on meter. 

Place the reference thermometer and your Testr 

into the same sample. Allow enough time for both to stabilize. 

2.

 

Press 

 as needed to select the pH measuring mode. Press 

 to begin pH 

calibration mode. 

3.

 

Press 

 for 5 seconds to begin temperature calibration mode. The current 

temperature will be displayed on top while the factory default temperature is 

below.  

4.

 

Press 

 or  

 to manually adjust to the desired temperature—up to ± 5° 

C or ± 9° F of the factory default value. 

5.

 

Press 

to confirm and return to the pH measuring mode.

 

pH Calibration: 

For best results, calibrate with certified accurate pH calibration standards (buffers). 

You may calibrate up to five points with the USA (1.68, 4.01, 7.00, 10.01, 12.45) or 
the NIST (1.68, 4.01, 6.86, 9.18, 12.45) buffer group. 

1.

 

Press 

 to turn meter on and 

 to select pH mode as needed. 

2.

 

Rinse the sensor with clean water. Immerse the sensor into your pH buffer and 

press 

. The primary display will show the un-calibrated pH value, while the 

secondary display should search for and lock on the closest automatic calibration 

value.  

3.

 

Allow the primary display to stabilize, then press 

 to confirm the calibration 

value. The primary value will blink briefly before the secondary value automatically 

scrolls thru the remaining pH buffers available for calibration.  

4. 

Repeat steps 2 & 3 with additional buffers or press 

 to return to 

measurement mode.

 

 

 

www.

GlobalTestSupply

.com

Find Quality Products Online at:

[email protected]

Reviews: