2230XP Silica Analyzer
Thermo Scientific Orion
60
8
Terms Definition
Terms
Description
Possible Values, Options
ALARM
Type
Assigning a fault type to the alarm.
Low Limit, High Limit, Fluidics Fault,
General Fault, not Assigned
Value
Boundary value for triggering an
alarm, only for low limit and high
limit type.
0 to XXXXX (XXXXX Value depends on
precision selected in Setup
➔
Display
parameters
➔
result Decimal Point)
Hysteresis
Value gap or difference before an
event change.
0 to XXXXX (XXXXX Value depends on
precision selected in Setup
➔
Display
parameters
➔
result Decimal Point
RELAY
Up to FOUR spare relays available.
MODE
Allows fault ON or OFF operation.
Normal or Fail Safe
ANALOG OUTPUT
Type
Normal or reverse acting.
0 to 20, 4–20, 20-0 or 20–4
Low
Low concentration value.
High
High concentration value.
EXTERNAL INPUT
Label
User programmable label.
Event
Dry contact event.
Close/Open or Open/Close
Action
Selected function.
Analyze Once, Run/Stop, Calibrate,
Validate, Not assigned
DISPLAY PARAMETERS
Contrast
To adjust the display contrast.
120 to 200
Back Light On-Time
Screen saver time duration for
backlight of display.
0 to 255 minutes, 0 for Continuously On