Section 6
Calibration
of the Instrument for Measurement Analysis
48
Thermo Scientific Orion
2230XP Silica Analyzer
Measurement Data
The Orion 2230XP Silica Analyzer product stores measurement and calibration results for up to
31 days. The data for each sample analysis cycle or calibration may be retrieved under
REVIEW RESULTS menu screen with the corresponding time stamp and other related
operating parameters.
Sample analysis data stores the date, time, concentration, units, temperature, wavelength and
status (Good or Fault) along with the faults for each measurement cycle.
Calibration data stores date, time, standard concentration, slope, intercept, temperature,
wavelength and status.
The stored results are displayed with respect to the selectable period of time such as: Last 12
Hours, Last 1 Day, Last 2 Days, Last 3 Days, Last 1 week, Last 2 weeks, Last 3 weeks, Last 4
weeks, All, which is selected as DATA RANGE. To display the measurement results from the
main menu follow the sequence:
MENU
➔
VIEW DATA AND ALARMS
➔
REVIEW ANALYSIS RESULTS
➔
DISPLAY
NOTE:
Validation results are ONLY reported on the display and the RS-232 at the time
of running the validation. Validation cycles are not stored and are not reported
on the analog outputs.
Printing Results
The stored measurement results can be printed on to the Hyper Terminal (minicom in Linux)
through the RS-232 port within the selected data range. The selected data range is applicable
to displaying, printing and plotting of measurement results.