Calibration of the Instrument for Measurement Analysis
Section 6
2230XP Silica Analyzer
Thermo Scientific Orion
43
If the STOP key is pressed during this measurement, it aborts the measurement and will restart
after the idle time countdown expires.
To stop the auto restart of the instrument the mode should be changed to manual restart:
MENU
➔
SETUP
➔
(+) ANALYSIS PARAMETERS
➔
MODE
Manual Restart Mode
This mode allows you to START the measurements manually.
To perform measurement on command from the MAIN MENU follow the sequence:
MENU
➔
MEASUREMENT
➔
ANALYZE ONCE
Pressing RUN key in this mode keeps the instrument in RUN MODE and pressing STOP key
during measurement gets back to the MANUAL RESTART mode. In this mode, if the instrument
is untouched for a period of the standby mode idle time (60 minutes default), it switches to
standby mode.
During standby operation, all fluidics devices are shut down and the analyzer goes into a
standby housekeeping mode in which the pumps and the valves are activated on an hourly
basis to prevent tube pinching.
When manual restart mode is selected, the analyzer automatically goes into standby operation
after one hour (60 minutes) of inactivity. Consult the factory should you require a different
standby time out; for details, refer to Section 7.
NOTE:
Even though the instrument is running continuously and displays RUN
MODE, to know the actual RESTART mode of the instrument from MAIN
MENU follow the sequence:
MENU
➔
SETUP
➔
(+) ANALYSIS PARAMETERS
➔
MODE