Orbitrap Tribrid Series Getting Connected Guide
15
5
Connecting the Cables and Power Cords
Read and understand the following cautions before you make the electrical connections in the
next procedure. If the API source contains the optional APPI lamp, you must provide a USB
splitter to accommodate the three USB connections. See also
Chapter 6, “External Devices.”
CAUTION Avoid an electric shock and equipment damage.
• Forepumps—Read the warning on the forepump. When you disconnect the
forepump’s power cord from the electrical outlet do not touch the plug’s pins for at
least 5 minutes to allow time for the voltage potential to dissipate.
• Mass spectrometer—Be aware of the following:
–
The AC Output receptacle is not used at this time—do not plug the forepump
(or any other appliance) power cord into this receptacle; see step 5 on
.
–
Before you begin the installation, place the Main Power switch in the Off (down)
position and the Electronics service switch in the Service Mode (down) position.
• Syringe pump—Make sure that the voltage selector switch is set to the appropriate
voltage (110 V or 220 V) for your country or territory.
CAUTION
Safety and EMC regulations require the use of Category 5e, shielded Ethernet
communication cables, maximum 3 m (10 ft) long.
CAUTION
After completing these connections, route all cables and power cords so that
they are not a trip hazard.