background image

 

Installation Kits and Power Cords

Dual Mechanical Pumps Kit

Thermo Scientific

Orbitrap Fusion Series Getting Connected Guide

23

Dual Mechanical Pumps Kit

Table 6.

Dual Mechanical Pumps Kit (P/N 80011-62006)

Image

Item

Quantity

Part number

Forepump, Oerlikon Leybold Vacuum SOGEVAC 
SV65BI (rated single-phase 230 Vac, 50/60 Hz)

2

00108-01-00032

Forepump oil tray, stainless steel, caster wheels

2

00201-99-00549

Dual relay control cable, 2.4 m (8 ft) long 
(preassembled) 

1

80100-63146

Dual pump vacuum hose assembly, including the 
stainless steel 45-degree elbows and connection parts 
(preassembled)

1

80011-60077

Exhaust waste connection

Fitting Tee, barbed, nylon, for 0.5 in. (12.7 mm) ID 
tubing

1

00103-01-00012

Tubing, Tygon, 3/4 in. (19.1 mm) OD, 0.5 in. 
(12.7 mm)  ID,  6 m  (20 ft)  long 

1

00301-22920

Summary of Contents for Orbitrap Fusion Lumos MS

Page 1: ...Orbitrap Fusion Series Getting Connected Guide 80011 97002 Revision C June 2017...

Page 2: ...ivision of Saint Gobain Performance Plastics Corporation Chemyx is a trademark of Chemyx Inc MX Series II is a trademark of IDEX Health Science LLC All other trademarks are the property of Thermo Fish...

Page 3: ...ement parts and additional components options and peripherals directly from Thermo Fisher Scientific or an authorized representative Regulatory compliance results for the following Thermo Scientific p...

Page 4: ...ve 2014 30 EU and other EMC test standards This device was tested by T V Rheinland of North America and complies with the following EMC standards 47 CFR 15 Subpart B Class A 2012 EN 61326 1 2013 EN 61...

Page 5: ...you purchased the product Number of product pieces and the estimated total weight and volume Pick up address and contact person include contact information Appropriate pick up time Declaration of dec...

Page 6: ......

Page 7: ...Connections 5 Chapter 3 API Source and Waste Container Connections 7 Installing the API Source 7 Connecting the Solvent Waste Container 8 Chapter 4 Syringe Pump and Modular Valve Assembly 11 Chapter 5...

Page 8: ...Contents viii Orbitrap Fusion Series Getting Connected Guide Thermo Scientific...

Page 9: ...tions such as the Thermo Xcalibur data system Also because this guide uses drawings of various connections and component parts to help illustrate procedures be sure to start from no matter where it ap...

Page 10: ...web browser save it or print it To return to the document list click the browser Back button Special Notices Symbols and Cautions Make sure you understand the special notices symbols and caution label...

Page 11: ...ispose of waste oil and when handling wetted parts of the instrument Heavy object The Orbitrap Fusion Series MS excluding its workbench weighs over 227 kg 500 lb Never try to detach and move the instr...

Page 12: ...o find global contact information or customize your request 1 Go to thermofisher com 2 Click Contact Us select the country and then select the type of support you need 3 At the prompt type the product...

Page 13: ...nd maintenance instructions do the following Prevent the forepump from overheating by ensuring that there is sufficient air clearance around the pump Maintain the exhaust pressure from atmospheric pre...

Page 14: ...r the MS 4 2 Using a centering ring and a swing clamp attach the plumbing elbow to the forepump s inlet port 3 Orbitrap Fusion Lumos MS Using a centering ring and a swing clamp connect a 90 degree plu...

Page 15: ...vel for at least 2 m 79 in This hose acts as a trap for exhaust fumes that would otherwise recondense in the forepump oil Although Thermo Fisher Scientific recommends that you periodically open the ga...

Page 16: ...1 Vacuum System Connecting the Forepumps to the Lab Exhaust System 4 Orbitrap Fusion Series Getting Connected Guide Thermo Scientific...

Page 17: ...ft3 Linear ion trap damping gas and collision gas Nitrogen UHP 99 999 Has less than 3 0 ppm each of water oxygen and total hydrocarbons 345 70 kPa 50 10 psi 46 L 1 62 ft3 Venting gas ion routing mult...

Page 18: ...another length of copper tubing except connect it between the Helium In inlet and your UHP helium gas supply 3 UHP nitrogen gas Cut a length of the copper tubing attach the 1 8 in ID Swagelok type nut...

Page 19: ...lvent Waste Container Installing the API Source All the wiring and gas plumbing for the API source are internal This means you can install or remove the API source or change the ionization mode1 H ESI...

Page 20: ...ntainer From the waste container to the exhaust system Equip your lab with at least two fume exhaust systems The analyzer optics become contaminated if the drain waste tubing and the exhaust tubing fr...

Page 21: ...e solvent waste container to the MS Connect the filling venting cap to the solvent waste container Then connect the Tygon waste tubing 1 in ID to the cap openings 1 Connect the tubing to the MS s Drai...

Page 22: ...3 API Source and Waste Container Connections Connecting the Solvent Waste Container 10 Orbitrap Fusion Series Getting Connected Guide Thermo Scientific...

Page 23: ...mp and plumb the modular valve refer to the Orbitrap Fusion Series Getting Started Guide Using the screwdriver and two of the provided screws attach the bottom shelf to the holder 1 Not shown Place th...

Page 24: ...4 Syringe Pump and Modular Valve Assembly 12 Orbitrap Fusion Series Getting Connected Guide Thermo Scientific...

Page 25: ...let do not touch the plug s pins for at least 5 minutes to allow time for the voltage potential to dissipate Mass spectrometer Be aware of the following The AC Output receptacle is not used at this ti...

Page 26: ...LED remains off press its front Remote button to exit local mode 5 1 Connect the relay control cable between the MS V65 Pump Enable and the forepumps 4 Not shown Connect the contact closure cable and...

Page 27: ...ck the front wheels by moving the locking levers forward to the upright position The locking levers have three positions To fully unlock the roll and swivel movements push the levers back to the third...

Page 28: ...5 Electrical Connections 16 Orbitrap Fusion Series Getting Connected Guide Thermo Scientific...

Page 29: ...try applications can control external devices for example autosamplers pumps and detectors from several manufacturers including Thermo Fisher Scientific Agilent Technologies and Waters Corporation Whe...

Page 30: ...e that the signal from the external device meets the requirement stated in the pin 4 Start In pin out description see Table 3 If it does not meet this requirement you cannot use it with the MS To help...

Page 31: ...d closes when the method finishes Output Maximum 24 Vdc 3 A 6 Injection Hold Provides a relay driven programmable output signal to the connected external device such as a fraction collector Output Max...

Page 32: ...e of the MS Other connectors USB 2 ports Provides a connection for the syringe pump and modular valve Ethernet 1000 Base T Provides a connection for the Ethernet switch Table 3 Pin out descriptions fo...

Page 33: ...Syringe Pump and Modular Valve Assembly Single Mechanical Pump Kit Dual Mechanical Pumps Kit MS Setup Kit Power Cords Syringe Pump and Modular Valve Assembly Table 4 Syringe pump and modular valve as...

Page 34: ...Forepump oil tray stainless steel caster wheels 1 00201 99 00549 Relay control cable single pump 2 4 m 8 ft long preassembled 1 80000 63139 Single pump vacuum hose assembly KF40 preassembled exploded...

Page 35: ...rated single phase 230 Vac 50 60 Hz 2 00108 01 00032 Forepump oil tray stainless steel caster wheels 2 00201 99 00549 Dual relay control cable 2 4 m 8 ft long preassembled 1 80100 63146 Dual pump vacu...

Page 36: ...5 port gigabit 1 00825 01 00111 Gas connections Ferrule brass back 1 4 in ID 2 00101 04000 Ferrule brass back 1 8 in ID 2 00101 02500 Ferrule brass front 1 4 in ID 2 00101 10000 Ferrule brass front 1...

Page 37: ...vided for Thermo Fisher Scientific field service use only Plug configuration Plug rating Mass spectrometer 2 5 m 8 ft North America P N 96000 98035 NEMA 6 15 250 Vac 15 A Internationalb P N 80000 6318...

Page 38: ...A Installation Kits and Power Cords Power Cords 26 Orbitrap Fusion Series Getting Connected Guide Thermo Scientific...

Page 39: ...2 forepumps fume exhaust system cautions 8 gas ballast valve opening 3 maintenance instructions 1 G gases specifications and consumptions 5 K kits installation 21 P power cords 14 25 procedures API so...

Page 40: ...28 Orbitrap Fusion Series Getting Connected Guide Thermo Scientific Index V V valve LED 14 viewing documents x W WEEE directive v...

Reviews: