4
Learning Center
Prepare Samples and Blanks
Thermo Scientific
NanoDrop One
C
with NanoDropQC Software User Guide
69
Run a Blanking Cycle
Run a blanking cycle to verify the following:
• instrument is operating normally (with flat baseline)
• pedestals are clean (i.e., no dried-down sample material on pedestals)
• absorbance contribution of buffer solution you plan to use for sample analyses
Supplies needed
• lint-free laboratory wipes
• calibrated precision pipettor (0–2 µL)
• buffer solution for evaluation
To run a blanking cycle
For quick demonstration, watch multimedia training
Evaluating a Blanking Solution for
Suitability
.
NOTICE
• Do not use a squirt or spray bottle on or near the instrument as liquids will flow into
the instrument and may cause permanent damage.
• Do not use hydrofluoric acid (HF) on the pedestals. Fluoride ions will permanently
damage the quartz fiber optic cables.