Maintenance
Instrument Diagnostics
130
NanoDrop One
C
with NanoDropQC Software User Guide
Thermo Scientific
Pedestal Image Check
Run the Pedestal Image Check periodically to verify the instrument’s column sensor which
monitors for possible errors such as an empty column or bubbles in a sample. The Pedestal
Image Check can be used for routine quality control purposes. It also provides important
diagnostic information if a detection system component fails.
Supplies needed
• lint-free laboratory wipes
To run pedestal image check
1. Lift the instrument arm and clean the upper and lower pedestal with a new laboratory
wipe.
2. Lower the arm.
3. From the instrument home screen, tap
(Diagnostics) and then tap
Pedestal Image Check
. If you are using the PC Control software, from the Home screen,
select
Pedestal Image
.
4. Tap
Measure
or select
Run.
The instrument runs a series of tests to check pedestal position and image quality. After
the measurements are completed, the results are displayed. A green check mark indicates
the instrument passed the Pedestal Image Check.
5. When finished, tap
End Experiment
.
To interpret pedestal image check results
If the Pedestal Image Check displays a yellow triangle instead of the green check mark, follow
the on-screen instructions to fix any possible problems. Then rerun the Pedestal Image Check.
If the instrument fails again,