![Thermo Scientific Multidrop Combi nL User Manual Download Page 68](http://html1.mh-extra.com/html/thermo-scientific/multidrop-combi-nl/multidrop-combi-nl_user-manual_1104640068.webp)
Maintenance
Regular and preventive maintenance
68
Thermo Scientific Multidrop Combi nL User Manual
Thermo Fisher Scientific
Procedure
D
aily
W
eekl
y
M
ont
hl
y
Year
ly
Replace the tubing set and a manifold, if necessary*. See “Replacing a
tubing set and a manifold” on page 77.
Ensure proper shutdown. See “Shutdown” on page 65.
Service the instrument regularly.
Decontaminate the instrument when relocating the instrument or
sending it for service. See “Decontamination procedure” on page 80.
Refit the transport lock when relocating the instrument or sending it for
service. See “How to refit the transport lock” on page 82.
= depending on the laboratory conditions and the use and configuration of the instrument
* = when necessary
For reliable daily operation, keep the instrument free of dust and liquid
spills. To prevent unnecessary wear or hazards, follow the routine and
service procedures described below at regular intervals.
Clean the outside of the instrument periodically with a cloth dampened
with water or a mild detergent (e.g. dishwashing liquid) or 70% ethanol
when necessary. Immediately wipe away spilled saline solutions,
solvents, acids or alkaline solutions from outer surfaces to prevent
damage.
Abrasive cleaning agents are not recommended, because they are likely
to damage the plastic instrument cover.
Caution
Always switch the instrument off and disconnect the power
cable before cleaning the dispenser.
s
Caution
The surfaces can be cleaned with most laboratory detergents.
Dilute the cleaning agent as recommended by the manufacturer. Do
not expose the surfaces to concentrated acids or concentrated alcohols
for prolonged periods of time as damage may occur.
s
It is recommended to service the instrument at least yearly. Refer to
“Service contracts” on page 84.
If you believe that liquid has entered the Multidrop Combi nL, first
switch the instrument off (Figure 2–4) and unplug the instrument.
Carry out corrective measures. Refer to “Emergency situations” on page
66 and “Decontamination procedure” on page 80 for aid. If necessary,
contact your local Thermo Fisher Scientific representative or the
Thermo Fisher Scientific technical service department. Refer to
“Packing for service” on page 83.
Regular and
preventive
maintenance
Cont.
Summary of Contents for Multidrop Combi nL
Page 1: ...Thermo Scientific Multidrop Combi nL User Manual Rev 1 2 Cat no N07171 ...
Page 100: ...Glossary 100 Thermo Scientific Multidrop Combi nL User Manual Thermo Fisher Scientific ...
Page 104: ...104 Thermo Scientific Multidrop Combi nL User Manual Thermo Fisher Scientific ...
Page 105: ...Thermo Fisher Scientific Thermo Scientific Multidrop Combi nL User Manual 105 Notes ...
Page 106: ...106 Thermo Scientific Multidrop Combi nL User Manual Thermo Fisher Scientific Notes ...