B-8
Micro-Tech 9101/9201 Reference Manual, Rev J
Thermo Fisher Scientific
6.
Press the Enter button to save the appropriate choice.
7.
Press the down-arrow button, and the Micro-Tech displays an
additional data-entry screen. Which screen is displayed depends on
which calibration (R-Cal, weights, or chain) method you chose
above. Please refer to the appropriate screen below..
R-Cal Method
Test-Chain Method
BATCH
- CALIB DATA SCROLL 1 -
Calibration mode
> Weights <
READY
ALARM
CALIB
ENTER
CHOICE
BATCH
- CALIB DATA SCROLL 1 -
Calibration mode
> Chain <
READY
ALARM
CALIB
ENTER
CHOICE
BATCH
- CALIB DATA SCROLL 2 -
R-Cal selected
Res 165000 Ohms
READY
ALARM
CALIB
EDIT
BATCH
- CALIB DATA SCROLL 4 -
Chain select. weight
0.000 Lb/Ft
READY
ALARM
CALIB
EDIT
Summary of Contents for Micro-Tech 9101