Maintenance and Troubleshooting
Opto22 Problems
4-16
Micro-Tech 9101/9201 Reference Manual, Rev J
Thermo Fisher Scientific
If GOOD—Proceed to “View Opto LED” (see step 4).
If BAD—Replace fuse.
4.
View Opto22 LED. By forcing the output open and closed, does the
LED indicate the appropriate logic condition?
If GOOD—Replace the Opto22 output module.
If BAD—Replace either the Opto22 output module or the Micro-
Tech CPU board.
Figure 4–10.
Opto22 Troubleshooting—Output Module