Tuning and Calibrating Automatically in the ESI/MS Mode
Tuning the MS Detector Automatically in the ESI/MS Mode
Thermo Electron Corporation
LTQ XL Getting Started
57
4. In the Mass spin box, enter
195.1
to specify that the LTQ XL MS
detector optimize your Tune Method on the peak at
m/z
195.1.
5. Start the automatic tuning procedure, as follows:
a. Click
Start
.
A message box displays the following message:
Please ensure that the 500 microliter syringe is full.
Ensure that the syringe contains at least 450
μ
L calibration solution.
b. Click
OK
to close the message box, and return to the Tune dialog
box.
6. On the File/Display toolbar, click the
Graph View
button to display the
Graph view. See
.
7. Observe the Tune Plus window and the Tune dialog box. While
automatic tuning is in progress, the LTQ XL MS detector displays
various tests in the Spectrum and Graph views in Tune Plus and displays
various messages in the Status group box in the Tune dialog box. Your
Tune Plus window should now look similar to the one shown in
8. Click the
ESI Source
dialog box to examine the ESI source parameters
after tuning. Compare the settings shown in
with the pre-tune
settings shown in Figure 19 on page 3-49.
9. Click the
Ion Optics
toolbar button to display the Ion Optics dialog
box. The parameters in the Ion Optics dialog box are optimized
automatically by the LTQ XL MS detector. See
.
You have now successfully tuned the MS detector in ESI/MS mode using
the calibration solution. Go to the next section,
.
Note
In this example, you use the mass peak at
m/z
195.1 to
optimize the Tune Method. However, you can optimize the tune on
any mass peak of the calibration solution.