Operation
Mode/Fault Alarm Setup
System Tab
To access the information on the Mode/Fault Alarm Setup System tab using only the
remote transmitter keypad, see
Figure 4-6.
Mode/Fault Alarm Setup, System Tab
Alarms have three operational modes.
–
Disable – The alarm is disabled, and no alarm action is executed.
–
Enable – The alarm is enabled. The selected alarm action is executed based on the
alarm set condition.
–
Inhibit – The alarm is temporary disabled. Another operation may change the
mode of this alarm from Inhibit to Enable.
The following alarms are available for configuration on the System tab of the
Mode/Fault Alarm screen.
1.
Hold Mode Active Alarm
a.
The alarm is triggered when any of the items listed below are in hold mode.
i.
Analog output A or B or C.
ii.
Relay A or B
iii.
Measurements 1, 2, 3 or 4.
b.
The alarm clears when all of the items listed above are out of hold mode.
2.
RBP Communication Failed Alarm:
a.
The alarm is triggered when an RBP communication errors occurs.
b.
The alarm clears when the RBP communication errors are cleared.
3.
Fieldbus Communication Failed Alarm
a.
This alarm has not yet been implemented, but will be available in future
releases.
Thermo Scientific
LevelPRO User Manual
4-9